Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
Table 2-1: Final Test and Calibration Values (Continued)
Parameter
Observed
Value
Units Nominal Range Reference Section
NO Span Conc PPB 20 - 20,000 Table 7-7-4
NO
x
Span Conc PPB 20 - 20,000 Table 7-7-4
NO Slope - 1.0 ± 0.3 Table 7-7-21
NO
x
Slope - 1.0 ± 0.3 Table 7-7-21
NO Offset mV -10 to +150 Table 7-7-21
NO
x
Offset mV -10 to +150 Table 7-7-21
Moly Efficiency % 0.96 - 1.02 7.8.6, 5.2.2.6
Stability at Zero PPB < 0.2 Table 9-9-1
Stability at Span PPB < 2 ppb @ 400ppb
span gas
Table 9-9-1
Measured Flows
Sample Flow cc/min 500 ± 50 9.3.7, Table 9-9-19
Ozone Flow cc/min 60 - 90 9.3.7, Table 9-9-19
IZS Purge Flow cc/min 60 ± 15 6.3
Factory Installed Options Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Rack Mount, External Pump w/ Slides
Rack Mount, External Pump w/o Slides
Stainless Zero/Span Valves
Internal Zero/Span - IZS
Permeation Tube Output Specification
Current Loop - NO
x
Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - NO Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - NO
2
Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Current Loop - TST Chan 4-20 mA 0-20 mA Isolated Non-Isolated
Internal Pump
PROM # Serial #
Date Technician
2-13