Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
7.2 Manual Zero/Span Check or Calibration with
Zero/Span Valves Option
The Zero/Span valve option can be operated from the front panel keyboard. In the Zero/Span
valve option the zero and span gas come into the valves through ports on the rear panel of the
instrument.
Table 7-7-6: Manual Zero Calibration Procedure - Z/S Valves
Step Number Action Comment
1. Press CALZ The analyzer enters the zero calibrate mode. This switches the
sample/cal and zero/span valves to allow zero gas to come in
through the zero gas inlet port in the rear panel.
2. Wait 10 min Wait for reading to stabilize at zero value.
3. Press ZERO If you change your mind after pressing ZERO, you can still
press EXIT here without zeroing the instrument.
4. Press ENTR Pressing ENTR actually changes the calculation equations,
forcing the reading to zero.
5. Press EXIT M200A returns to sample mode. Immediately after
calibration, readings do not go into the DAS averages.
Refer to Table 7-7-4 to enter expected NO and NO
x
values.
Table 7-7-7: Manual Span Calibration Procedure - Z/S Valves
Step Number Action Comment
1. Press CALS The M200A enters the calibrate mode from sample mode.
This operates the sample/cal and zero/span valves to allow
span gas to come in through the cal gas inlet port in the rear
panel or optional permeation tube.
2. Wait 10 min Wait for reading to stabilize at span value.
3. Press SPAN If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.
4. Press ENTR Pressing ENTR actually changes the calculation equations.
5. Press EXIT M200A returns to sampling. Immediately after calibration,
data is not added to the DAS averages.
7-6
PRINTED DOCUMENTS ARE UNCONTROLLED