The T command lists TEST messages. Examples of the T command are:
T LIST
Lists all active test messages
T LIST ALL
Lists all test messages
T CONVTEMP
Prints the temperature of the moly converter
T LIST NOX
Prints NO
x
concentration message
T NOX
Prints NO
x
concentration message
5.5.4 WARNING Commands and Messages
Table 5-5-18: RS-232 Warning Messages
Name Message Description
WSYSRES SYSTEM RESET Analyzer was reset/powered on
WRAMINIT RAM INITIALIZED RAM was erased
WSAMPFLOW SAMPLE FLOW WARN Sample flow out of spec.
WOZONEFLOW OZONE FLOW WARNING Ozone flow out of spec.
WRCELLPRESS RCELL PRESS WARN Rx cell pressure out of spec
WBOXTEMP BOX TEMP WARNING Box temp. out of spec.
WRCELLTEMP RCELL TEMP WARNING Reaction cell temp. out of spec.
WIZSTMP IZS TEMP WARNING IZS temp. out of spec.
WCONVTEMP MOLY TEMP WARNING Molycon temp. out of spec.
WPMTTEMP PMT TEMP WARNING Molycon temp. out of spec.
WAUTOZERO AZERO WRN XXX.X MV AutoZero filter received a reading out
of limit spec.
WHVPS HVPS WARNING High voltage out of spec.
WDCPS DCPS WARNING DC Voltage out of spec.
WOZONEGEN OZONE GEN OFF Ozone generator is off
WDYNZERO CANNOT DYN ZERO Dynamic zero cal. out of spec.
WDYNSPAN CANNOT DYN SPAN Dynamic span cal. out of spec.
WVFDET V/F NOT DETECTED V/F board not installed or broken