ESD Precautions
The CCD and other system electronics are extremely sensitive to electrostatic discharge
(ESD). To avoid permanently damaging the system, please observe the following
precautions:
When using high-voltage equipment (such as an arc lamp) with your I-PentaMAX
system, be sure to turn the controller power on last and power the controller off
first.
Use caution when triggering high-current switching devices (such as an arc
lamp) near your system. Transient voltage spikes can permanently damage the
CCD. If electrically noisy devices are present, an isolated, conditioned power
line or dedicated isolation transformer is highly recommended.
Never connect or disconnect any cable while the I-PentaMAX system is powered
on. An unconnected cable segment can become electrically charged and can
damage the CCD if reconnected.
Connect the camera-power supply cable to the Temperature/Power Supply unit
before connecting the cable to the camera.
Disconnect the camera-power supply cable from the camera before
disconnecting it from the Temperature/Power Supply unit.
Additional Precautions
Camera and Temperature/Power Supply Unit
Never impede airflow through the equipment by obstructing the air vents. Allow
at least a one-inch air space around any vent.
Do not “mix and match” cameras and Temperature/Power Supply units.
Image Intensifier Controller (IIC-200, IIC-300, and IIC-100)
Before turning on the high voltage supply to the intensifier, turn MCP GAIN to 0
gain.
Environmental Requirements
Storage temperature -20°C to 55°C
Operating environment 0°C to 30°C
Relative humidity 50%.