Princeton Digital (USA) 4411-0062 Universal Remote User Manual


 
94 I-PentaMAX System Manual Version 3.A
REGION msec/FRAME FPS*
160 × 160 7.07 141
98 × 98 3.08 324
89 × 89 2.72 367
68 × 68 1.74 574
56 × 56 1.39 719
51 × 51 1.14 877
41 × 41 0.92 1087
38 × 38 0.78 1282
32 × 32 0.68 1470
32 binned × 512 0.52 1923
Table 9. I-PentMAX, 5 MHz: Virtual Chip Size
and Approximate Number of Frames per Second
* Virtual Chip speeds determined from scan code
calculator assuming exposure equals readout time.
Applies only to I-PentaMAX with 5 MHz ADC
Virtual Chip Setup
Introduction
If the Virtual Chip mode option has been installed, both WinView/32 and WinSpec/32
will support this technique. The following procedure covers the basic hardware and
software setup for Virtual Chip operation.
Note: The Virtual Chip dialog box is discussed in detail in the next section.
Equipment:
I-PentaMAX (Version 5) with 512x512FT CCD array
IIC-200 and High Voltage cable
Temperature/Power Supply unit and Camera to Power Supply cable
Princeton Instruments (RSPI) PCI Interface Card and High Speed Serial (TAXI)
cable
75 BNC cable
Suitable Host Computer
Software:
WinView/32, version 2.4 or higher
WXvchip.opt installed in the same directory as the executable WinView/32 program