Changing the meter size of the detector
The following is an example to change the meter size from 50mm to 10mm.
Switch operation Display example Description
A2: METER SIZE
50 mm
Select A2: METER SIZE from the setting item selection.
The currently set meter size (50mm in this example) appears.
A2: METER SIZE
50
mm
The cursor appears.
A2: METER SIZE
100
mm
Change the set value for meter size to 100mm.
The set value for meter size can be changed using
switch.
(Note 1)
A2: METER SIZE
100 mm
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
A
2: METER SIZE
Returns to the menu of setting item selection.
Note 1: The meter size is displayed cyclically as shown below:
* For partially filled pipes detectors (LF232*F) : 6
24inch (150600mm )
Note 2: If the setting value for meter value is changed, the setting value by flow unit and the setting
value for exciting frequency of no concern may be rewritten.
After the meter size is set, be sure to check the setting values of Range, Count (Pulse) Rate
and Exciting Frequency.
In the case of the converter for large meter
size detector (LF232*B)
600mm2.5mm 700mm 3000mm
0.1inch
24inch
28inch 120inch