Agilent Technologies E4400-90246 TV Converter Box User Manual


 
104 Chapter 3
PSA Functional Tests
Amplitude Accuracy at 50 MHz
PSA Functional Tests
7. On the spectrum analyzer, press:
BW/Avg, Average (On)
20,
Enter
8. Perform the following steps for each row listed in Table 3-1:
a. Set the synthesized sweeper amplitude to the value listed in the Nominal Source
Amplitude column in Table 3-1.
b. Set the Attenuation and Span as listed in each row of the table.
c. Record the source amplitude, as measured by the power meter, in the Power Meter
Amplitude column of Table 3-1.
d. On the spectrum analyzer, press
Restart, Peak Search.
e. Wait for the spectrum analyzer to finish averaging.
f. Record the signal amplitude, as measured by the analyzer in the Measured
Amplitude column of Table 3-1.
g. Calculate the signal amplitude accuracy error using the following equation, and
record the results under the Amplitude Accuracy Error column:
Amplitude Accuracy Error = Meas_amp
Power_meter