Chapter 1 53
Instrument Messages
Instrument Messages
Instrument Messages
769 ESA only Invalid instrument mode
You have attempted to switch to an instrument mode that is currently not
installed. Confirm that the mode name (for INST:SEL) or number (for
INST:NSEL) was entered correctly and that the requested personality is
actually installed in the instrument.
770 Incorrect filename, allowable extensions are .gif or .wmf
You sent a remote command to save a screen file but did not specify a valid
extension.
Incorrect filename, allowable extension is .sta
You sent a remote command to save a state file but did not specify a valid
extension.
Incorrect filename, allowable extensions are .trc or .csv
You sent a remote command to save a trace file but did not specify a valid
extension.
770 Instrument mode requested is not supported
Instrument mode specified with: INST command is not valid. Refer to
Chapter 5, Instrument Subsystem of Agilent Technologies ESA Series
Spectrum Analyzer Programmers Guide for more information.
771 Store Ref trace before turning on Normalize
A reference trace must be available for the Normalize function to be
activated. Refer to View/Trace in the Agilent ESA Spectrum Analyzer
Users Guide where the
Normalize key function is explained in detail.
772 Cannot load a directory
Cannot load a directory, please choose a file
You have selected a directory instead of a file when attempting to perform
the Load function under the
File front-panel key.
780 No Peak Found
781 Video Trigger cannot be active with Average Detector
You sent a remote command to do one of the following:
Turn on video trigger while the Average Detector or a Marker Function
is active.
Turn on the Average Detector or a Marker Function while Video Trigger
is active.
782 Video Trigger cannot be used with Mkr Func due to Average
Detector
783 Allowable span for current center frequency exceeded
784 Allowable CF for current span exceeded