Instrument Messages
Instrument Messages
Instrument Messages
10413 B7D DSP Code Install Fail Opt.229. Contact your Service
Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are not present in the analyzer. Reinstall the measurement
personality and contact your service center.
10414 Opt 229 DSP algorithm code file requires upgrade.
The Digital Signal Processing algorithm code file revision is not supported
by the currently loaded personality version. Refer to the users guide for
the personality in use for more information on installation/upgrade.
10415 Opt 229 DSP algorithm coef. file requires upgrade.
The Digital Signal Processor algorithm coefficient file revision is not
supported by the currently loaded personality version. Refer to the users
guide for the personality in use for more information on
installation/upgrade.
10416 B7D DSP Code Reload Fail Opt.229. Contact your Service Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are corrupt and cannot be loaded properly. Reinstall the
measurement personality and contact your service center.
10417 This operation requires a measurement to be active.
The analyzer cannot perform this operation, because it requires a
measurement to be running and no measurement is running.
10418 Demod Format setting does not allow this value for
Points/Symbol.
Invalid Points/Symbol selection attempted for the given Demod Format
setting.
10419 Carrier Not Present.
A carrier signal/burst is expected at the analyzer input. This signal cannot
be found; however, the measurement will still run. Check input signal
connection.
10420 Valid burst not found.
This error only occurs when
RF Amptd is selected for Burst Sync. It
indicates that a valid burst envelope which meets the specified Burst
Search Threshold was not found within the specified Burst Search Length.
Note that the demodulation measurement will proceed by attempting to
demodulate the signal without Burst Sync alignment. Possible causes may
be:
1. Carrier signal is not actually bursted.
2. Burst Search Threshold and/or Burst Search Length may need to be
adjusted.