Instrument Messages
Instrument Messages
Instrument Messages
253 Corrupt media
A removable media was found to be bad or incorrectly formatted. Any
existing data on the media may have been lost.
250 Mass storage error; EEPROM write timeout on EEPROM
Failure to initialize EEPROM. Report this error to the nearest Agilent
Technologies sales or service office.
241 Hardware missing
Hardware missing; no such SIO address
Missing device hardware. For example, an option is not installed.
230 Data corrupt or stale
Data corrupt or stale; EEPROM copy of <file>
Data corrupt or stale; NRAM copy of <file>
Possibly invalid data. A new measurement was started but not completed.
224 Illegal parameter value
You have sent a parameter for this command that is not allowed. See the
Function Reference.
223 Too much data; not all points entered
Too much data; <description of the type of data exceeded>
Indicates that a legal program data element of block, expression, or string
type was received that contained more data than the device could handle
due to memory or related device-specific requirements.
222 Data out of range
A legal program data element was parsed but could not be executed
because the interpreted value was outside the legal range defined by the
device (see IEEE 488.2 11.5.1.1.5). The displayed results may be clipped.
221 Settings conflict;
Averaging unavailable when Sig ID is on
User in ext mix with Sig ID on and try to select Averaging
Mode.
Cannot center unpreselected external mixer
Mixer Type is Unpreselected and user tries Preselector
Centering.
Command incompatible with band pair marker
Using remote commands, you have tried to adjust the start
or stop frequency of a span pair marker. You can adjust
only the center and span.