Instrument Messages
10450 Avg Mode changed to Repeat for Full Meas Type.
Meas Type has been changed to Full. Exponential Average Mode is not
available for Full Meas Type therefore Average Mode has been changed to
Repeat.
10451 Avg Mode changed to Exp for Examine Meas Type.
Meas Type has been changed to Examine. Repeat Average Mode is not
available for Examine Meas Type therefore Average Mode has been
changed to Exponential.
10452 Meas Type changed to Examine for Exp Avg Mode.
Average Mode has been changed to Exponential. Full Meas Type is not
available for Exponential Average Mode therefore Meas Type has been
changed to Examine.
10453 Meas Type changed to Full for Repeat Avg Mode.
Average Mode has been changed to Repeat. Examine Meas Type is not
available for Repeat Average Mode therefore Meas Type has been changed
to Full.
10454 Valid burst not found.
Cannot find a valid burst, either because it is not fully on the display, or
the level is too low to be detected properly.
10509 Did not find 4 signals > Peak Excursion.
or
Require 4 Signals > PEAK EXCURSION above Threshold.
Using the current instrument settings, the measurement cannot locate
four signals with enough energy or resolution to perform an accurate
measurement.
10510 Signals to not fit expected Intermod (TOI) pattern.
Based on the current analyzer settings, the measurement cannot find two
signals of high enough amplitude that can be resolved in such a manner
that the third order products would lie within the measured span.
10511 100 spurs found. Additional spurs ignored.
There are too many spurs for the table (the limit is 100), and any
additional spurs that are found will be ignored
10512 No spurs have been found.
You have started a measurement in examine meas type, in single or
continual sweep mode, or full meas type in single sweep mode, but no
spurs were found.
10513 No ranges are defined.
There are no active ranges in the range table. The user will need to
activate at least one range.