PSA Functional Tests
Frequency Response (Flatness)
PSA Functional Tests
Frequency Response (Flatness)
Test Limits
Test Description
The frequency response test measures the spectrum analyzers amplitude error as a
function of the tuned frequency. Measurements are made ranging from 50 MHz to the
maximum frequency range of your analyzer. The signal source amplitude is measured with
a power meter to eliminate error due to source flatness. The measured value is normalized
to 50 MHz.
Frequency Range
Limit Relative to
50 MHz
3 Hz to 3 GHz
±1.5 dB
> 3 GHz to 6.6 GHz ±2.5 dB
> 6.6 GHz to 22 Ghz ±3.0 dB
> 22 GHz to 26.5 GHz ±3.5 dB
> 26.5 GHz ±3.5 dB
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Adapter Type N(m) to 3.5 mm(m) 1250-1743
Adapter Type N(m) to 3.5 mm(f) 1250-1744
Adapter
a
2.4 mm(f) to 3.5 mm(m) 11901D
Adapter 3.5 mm(f) to 3.5 mm(f) 5061-5311
Cable 3.5 mm, 1 meter 8120-4921
Cables BNC, 120 cm (48 in.) 10503A
Power Meter Compatible with power sensor E4418B
Power Sensor Frequency Range: 50 MHz to 26.5 GHz 8485A
Power Sensor
a
Frequency Range: 50 MHz to 50 GHz 8487D
Power Splitter Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
11667B