154 Chapter 4
ESA Functional Tests
Scale Fidelity
ESA Functional Tests
Scale Fidelity
Test Limits
Test Description
A 50 MHz CW signal is applied from a synthesized sweeper to the input of the analyzer.
The source is adjusted for a response at the reference level. The synthesized sweeper
amplitude is adjusted to achieve a nominal amplitude below the reference level. The
analyzers amplitude marker is compared to the actual source change to determine the
scale fidelity error. Most of the error is the sources output attenuator inaccuracy from the
synthesized sweeper.
dB from Reference Level Minimum (dB) Maximum (dB)
4 1.0 1.0
16 1.4 1.4
28 1.4 1.4
40 1.4 1.4
52 1.4 1.4
64 2.0 2.0
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Signal Sources
Synthesized Sweeper Output Level Accuracy
0 to 15 dBm: ±1.0 dB
16 dBm to 63 dBm: ±1.4 dB
64 dBm:2.0 dB
8340A/B or 836XX
Series
Adapters
Type-N (m), to
BNC (f)
1250-1476
Cables
Type-N, 152-cm
(60-in)
11500D
BNC, 122-cm (48-in) 10503A
Additional Equipment for 75-Ohm Input
Pad, minimum loss 11852B
Type-N (f), to
BNC (m)
1250-1534