66 Chapter 1
Instrument Messages
Instrument Messages
Instrument Messages
10401 Valid burst not found in frame (Burst Type).
No active bursts that match the Burst Type have been detected in the RF
input signal. The search was performed over the complete frame.
10402 Valid burst not found in specified timeslot (Burst Type).
No active bursts that match the Burst Type have been detected in the RF
input signal. The search was only performed over the specified timeslot
setting.
10403 Sync word not found in frame (TSC).
One or more active EDGE bursts that match the Normal Burst Type have
been detected in the RF Input signal, but none contain the selected
Training Sequence Code (TSC). The search was performed over the
complete EDGE frame.
10404 Sync word not found in specified timeslot (TSC).
One or more active EDGE bursts that match the Normal Burst Type have
been detected in the RF Input signal, but none contain the selected
Training Sequence Code (TSC). The search was only performed over the
specified timeslot setting.
10409 Wideband calibration not valid until Align Now, All performed.
Wideband calibration corrections cannot be applied to current
demodulation results. Results will not meet specified demodulation
performance levels until an Align Now, All has been performed.
10410 Measurement uncalibrated, wideband calibration required.
Wideband calibration corrections cannot be applied to the current
demodulation results. Results will not meet specified demodulation
performance levels until
Wideband Cal has been selected and successfully
performed.
10411 Measurement uncalibrated, symbol rate exceeds +/-10% nominal.
Wideband calibration corrections cannot be applied to current
demodulation results, due to the
Symbol Rate setting exceeding +/-10% of
the nominal value. Results may not meet specified demodulation
performance levels.
10412 Cal Data corrupt. Wideband cal failed. Using previous data.
Wideband calibration procedure failed for an unknown reason. New
calibration data was not saved. Previous wideband calibration data is
being used, if available. However, it is not acceptable. Before re-attempting
Wideband Cal, ensure connection between the
AMPTD REF OUT and the
INPUT 50 is in place. If the error persists, get in touch with your service
center.