Instrument Messages
Instrument Messages
Instrument Messages
736 Too many data values at <freq_or_time_value>
This message may appear when data is sent to a corrections or limit table
using the
DATA or MERGE commands. These tables limit the number of
amplitudes associated with a frequency or time to 2 or less. This message
will appear if an attempt is made to attach 3 or more values to a frequency
or time.
751 Instrument state set to initial values
Instrument state may be corrupt, state has been reset to
initial values
While trying to load a trace or state, the state information was found to be
in error. This may be because the state had been stored on a later revision
of analyzer firmware. A default set of state variables was loaded instead.
There is nothing wrong with the analyzer.
An error in the internal instrument state has been detected. The state has
been reset to a default value.
752 Unable to load state from file
Attempt to load a state from a file failed.
753 Unable to save state to file
Attempt to save a state to a file failed. See the associated error messages
for the cause (press
System, Show Errors).
For PSA: If this error occurs after a mode switch that takes longer then
normal, you may need to free up some instrument memory. When you
switch to a different measurement mode, the analyzer must temporarily
store instrument state data. If you have less then 500 kB of available
memory, the switching process can take over 1 minute and result in this
error. Use the File key menu to delete any unnecessary files from
instrument memory.
755 Unable to load state from register
Attempt to load a state from an internal state register failed.
756 Unable to save state to register
Attempt to save a state to an internal state register failed. See the
associated error messages for the cause (press
System, Show Errors).
757 Unable to load user state, factory preset was done
A user preset failed, so the factory preset values were used. Save a valid
state into user preset and try again.
758 Unable to save user state
Attempt to save a user preset state failed. See the associated error
messages for the cause (press
System, Show Errors).