Agilent Technologies E4400-90246 TV Converter Box User Manual


 
Chapter 2 97
Functional Tests
What You Will Find in This Chapter
Functional Tests
Frequency Readout Accuracy on page 139
Marker Count Accuracy on page 141
Frequency Response (Flatness) on page 142
Reference Level Accuracy on page 146
Resolution Bandwidth Switching Uncertainty on page 151
Scale Fidelity on page 154
Second Harmonic Spurious Responses on page 157
Tracking Generator Level Flatness: Models E4401B and E4411B, Options 1DN and
1DQ on page 160
Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and
E4408B, Option 1DN on page 163
Each functional test includes:
Test limits (pass/fail criteria)
A description of the test
The equipment required for the test
A figure showing how to connect the equipment
Step-by-step instructions
One or more tables in which to record the measurement results
Before Performing a Functional Test
1. Ensure that you have the proper test equipment.
2. Switch on the unit under test (UUT) and let it warm up (in accordance with warm-up
requirements in the instrument specifications).
3. Allow sufficient warm-up time for the required test equipment (refer to individual
instrument documentation for warm-up specifications).
4. Ensure that the analyzers frequency reference is set to Internal:
a. Press the
System, Reference keys.
b. If the
Freq Ref softkey does not have Int underlined, press the Freq Ref softkey until it
is underlined.
5. Following instrument warm-up, perform the auto align routine:
Press
System, Alignments, Align All Now.
NOTE Functional test accuracy depends on the precision of the test equipment used.
Ensure that all of the test equipment is calibrated before running a functional
test.