Agilent Technologies E4400-90246 TV Converter Box User Manual


 
156 Chapter 4
ESA Functional Tests
Scale Fidelity
ESA Functional Tests
6. Record the marker delta reading in Table 4-7. At each new synthesized sweeper power
level, press
Single, Peak Search, and record the marker amplitude level.
Table 4-7 Scale Fidelity Worksheet
Synthesized
Sweeper Level
Minimum (dB)
Marker
Level (dB
)
Maximum (dB)
Reference=______ 0 (Reference) 0 (Reference)
Reference 4 dB 5.0 3.0
Reference 16 dB 17.40 15.60
Reference 28 dB 29.40 26.60
Reference 40 dB 41.40 38.60
Reference 52 dB 53.40 50.60
Reference 64 dB 66.0 62.0