Agilent Technologies E4400-90246 TV Converter Box User Manual


 
Chapter 1 55
Instrument Messages
Instrument Messages
Instrument Messages
10008 to 10517: Power Suite Error Messages An error number in this range indicates
the instrument has detected an error relating to the Power Suite functionality.
10008 Preferred resolution bandwidth not available.
The calculated required resolution bandwidth for this measurement is not
available.
10010 One or more harmonics past freq limit: number decreased.
Highest harmonic was past the frequency limit of the analyzer, so the
number of measured harmonics was decreased.
10011 First harmonic is past analyzer frequency limit.
First harmonic was past the frequency limit, so no harmonics can
be measured. The analyzer will simply measure the fundamental
. If
you have the Agilent E4407B analyzer, Option AYZ allows the use of
HP/Agilent 11970 Series, and HP/Agilent 11974 external mixers to extend
the frequency range to 110 GHz. Operation to 325 GHz is also possible
with non-HP/Agilent mixers.
10012 No Fundamental > 0 Hz found in given span.
There were no frequencies greater than 0 Hz in the starting
span, so the centre frequency at measurement entry will be used
as the fundamental
.
10013 No Fundamental > -50 dBm found in given span.
There was no fundamental found, so the centre frequency at measurement
entry will be used as the fundamental.
10020 Signal Tracking is not available when Noise Correction is on.
If signal tracking is turned on while noise correction is on, it will be turned
off. To use the signal tracking function, first turn noise correction off.
10138 Valid GSM burst not found in frame (Burst Type)
No active GSM bursts that match the selected burst type have been
detected in the RF input signal. Search was performed over a complete
GSM frame.
10139 Valid GSM burst not found in specified timeslot (Burst Type)
No active GSM bursts that match the selected burst type have been
detected in the RF input signal. Search was performed over the specified
timeslot.
10140 Valid GSM burst not found in frame (Ref Burst)
No active GSM bursts that match the selected burst type have been
detected in the RF input signal. Search was performed over the specified
complete GSM frame using the Ref Burst type setting.