Chapter 1 43
Instrument Messages
Instrument Messages
Instrument Messages
206 Unable to initialize flatness data
A failure occurred in setting the flatness data in the internal EEROM. Get
in touch with your local Agilent Technologies sales and service office.
207 Unable to store flatness data
A failure occurred in setting the flatness data in the internal EEROM. Get
in touch with your local Agilent Technologies sales and service office.
208 Preselector is not available at this frequency.
Preselection of frequencies below
3 GHz is not supported. Preselector
centering can be performed only for frequencies above 3.025 GHz.
209 Preselector centering failed
An attempt to center the preselector failed. You tried to center the
preselector with the analyzer span too wide. Preselector centering can be
performed only for spans below 1.123 GHz.
211 RBW limited to 1 kHz when Span > 5 MHz
In spans greater than 5 MHz, narrow (digital) resolution bandwidths,
below 1 kHz, are not available.
213 Span limited to 5 MHz when RBW < 1 kHz
In narrow (digital) resolution bandwidths, below 1 kHz, spans greater
than 5 MHz are not available.
214 TG start freq is less than 9 kHz
Tracking generator uncalibrated below 9 kHz.
215 TG start freq is less than 1/2 res bw
Tracking generator uncalibrated at start frequencies below 1/2 the current
resolution bandwidth.
216 Invalid Baud Rate
Attempt to use invalid baud rate. Refer to the programming language
chapter of Agilent Technologies ESA Series Spectrum Analyzer
Programmer’s Guide for more information.
217 RS-232 Interface Error
An error occurred on the serial interface.
219 Command not valid in this model
Indicates that the command sent from the remote interface does not apply
to this analyzer model number. For example, attempting to center the
preselector in an analyzer without a preselector will generate this error.
220 Trace Menu is unavailable when Sig Ident is ON