HP (Hewlett-Packard) 1660 TV Converter Box User Manual


 
Change the Oscilloscope Trigger
You will be using the timing analyzer to find the glitch and then the
oscilloscope to display it. This correlation of timing and scope is done by
arming the scope with the timing analyzer. To capture the glitch on the
scope, you want it to capture data immediately after the glitch is captured by
the timing analyzer.
1
Press the Trigger MENU key.
2 Highlight the Mode/Arm Edge field at the left side of the display, and
press the
Select key.
3 Highlight Immediate in the pop-up menu, and press the Select key.
Changing the Oscilloscope Trigger
Result The oscilloscope is set to capture data after the timing analyzer triggers on
the glitch.
Setting this field to
Immediate tells the
oscilloscope to trigger
immediately after the
timing analyzer triggers.
Triggering the Oscilloscope with the Timing Analyzer
Change the Oscilloscope Trigger
8-7