Tektronix OA 5000 TV Cables User Manual


 
Commands
OA 5000 Series User Manual
3 --- 15
Table 3-5: OA 5000 Device Commands and Parameters (Cont.)
Header Ful l Command Name
REFerence Reference
STORe1|2 Store Attenuation
WAVelength Wavelength
ADJusting? (Query Only)
The ADJusting? query returns the status of the attenuator. A 1 is returned if
the attenuator is moving to some attenuation value. A 0 is returned if the
attenuator is stationary.
*OPC, *WAI.
ADJusting?
ADJusting ?
ADJ?
would return the string ”:ADJUSTING 0” or ”:ADJUSTING 1”.
ALLev? (Query Only)
The ALLev? query causes the OA 5000 to return all events and their mes-
sages. This query also removes the returned events from the Event Queue.
The messages are separated by commas. Use the *ESR? query to enable
the events to be returned. For a complete discussion of the use of these
registers, see page 3---35. This command is similar to repeatedly sending
EVMsg? queries to the OA 5000.
*CLS, DESE, *ESE, *ESR?, EVENT?, EVMsg?, EV QTY, *SRE, *STB?
ALLev?
ALLev ?
The event code and message in the following format:
<Event Code><Comma><QString>[<Event Code><Com-
ma><QString>...]
Related Commands:
Syntax:
Examples:
Related Commands:
Syntax:
Returns: