OA 5000 Series User Manual A --- 3
Appendix B: Specifications
The performance characteristics described in this appendix (Tables A-1
through A-3) are valid under the following conditions:
H The instrument must have been calibrated at an ambient temperature
between +20_ C and +30_ C.
H The instrument must have had a warm-up period of at least 20 minutes.
H All optical connections to the instrument are made using physical con-
tact connectors. Tektronix offers limited selection of fiber optic jumper
cables which utilize physical contact on the connector ends (for exam-
ple: Tektronix part number 174-1387-00, see Appendix A).
H Both input and output connectors have been cleaned as described on
page A---15.
Table A -1: Optical Characteristics
Characteristic
Specification
Wavelength Range OA 5002 1100 --- 1600 nm
OA 5012/22/32 750 --- 1600 nm
Input
OA 5002
OA 5012
OA 5022
OA 5032
Single Mode Fiber
50 m Multimode Fiber
62.5 m Multimode Fiber
100 m Multimode Fiber
Insertion Loss
750 --- 1100 nm
1100---1350 nm
1350---1600 nm
≤ 4.0 dB (OA 5012/22/32 only)
≤ 2.0 dB
≤ 2.5 dB
Return Loss
OA 5002
OA 5012/5022/5032
≤ --- 45 dB
≤ --- 25 dB
Attenuation Range
OA 5002
1100 --- 1350 nm
1350---1600 nm
OA 5012/22/32
750---1350 nm
1350---1600 nm
≥ 60 dB (calibrated, 0.01 dB step size)
≥ 50 dB (calibrated, 0.01 dB step size)
≥ 60 dB (calibrated, 0.01 dB step size)
≥ 50 dB (calibrated, 0.01 dB step size)
Maximum Attenuation
OA 5002/5012/5022
OA 5032
≥ 100 dB (shutter activated)
≥ 90 dB (shutter activated)
Repeatability ±0.05dB