Agilent Technologies N3280A TV Cables User Manual


 
6 – Language Dictionary
58
CALibrate:DATA
This command enters a calibration value that you obtain by reading an external meter. You must first
select a calibration level (with CALibrate:LEVel) for the value being entered.
Command Syntax
CALibrate:DATA <NRf>
Parameters
<external reading>
Unit
A or V (amperes or volts)
Examples
CAL:DATA 3222.3 MA CAL:DATA 5.000
Related Commands
CAL:STAT CAL:LEV
CALibrate:DATE
This command stores the date the unit was last calibrated. Enter any ASCII string up to 31 characters.
Command Syntax
CALibrate:DATE <date>
Parameters
<date>
Examples
CAL:DATE "3/22/01" CAL:DATE "22.3.99"
Query Syntax
CALibrate:DATE?
Returned Parameters
<SRD>
CALibrate:LEVel
This command selects the next point in the calibration sequence. P1 is the first calibration point,
P2 is the second calibration point.
Command Syntax
CALibrate:LEVel <point>
Parameters
P1 | P2
Examples
CAL:LEV P2
CALibrate:PASSword
This command lets you change the calibration password. A new password is automatically stored in
nonvolatile memory and does not have to be stored with CALibrate:SAVE. The default password is the
model number of the unit. If the password is set to 0, password protection is removed and the ability to
enter the calibration mode is unrestricted.
Command Syntax
CALibrate:PASSword <NRf>
Parameters
<model number> (default)
Examples
CAL:PASS 1234
Related Commands
CAL:SAV