Agilent Technologies N3280A TV Cables User Manual


 
6 – Language Dictionary
62
SENSe:FUNCtion
This command configures the sensing function for triggered measurements. The dc source has two
measurement sensors as described below. The query returns the function setting.
CURRent
Senses the output current at the selected output
VOLTage
Senses the output voltage at the selected output
Command Syntax
SENSe:FUNCtion <function>, (@<channel list>)
Parameters
"VOLTage" | "CURRent"
*RST Value
VOLT
Examples
SENS:FUNC "VOLT",(@1)
Query Syntax
SENSe:FUNCtion? (@<channel list>)
Returned Parameters
<SRD>
SENSe:SWEep:NPLCycles
This command specifies the total measurement acquisition time in terms of ac power line cycles. It
automatically sets the sweep time interval, sweep offset, and sweep points. The values are chosen to
sample the maximum number of points possible and to provide the best noise filtering.
Command Syntax
SENSe:SWEep:NPLCycles <NRf+>
Parameters
1 through <n>
*RST Value
0.00912 (for 60Hz line)
0.0076 (for 50Hz line)
Examples
SENS:SWE:NPLC 10
Query Syntax
SENSe:SWEep:NPLCycles?
Returned Parameters
<NR3>
SENSe:SWEep:OFFSet:POINts
This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger. Positive values represent the delay after the trigger
occurs but before the samples are acquired.
Command Syntax
SENSe:SWEep:OFFSet:POINts <NRf+>
Parameters
-4095 through 2,000,000,000
*RST Value
0
Examples
SENS:SWE:OFFS:POIN -2047
Query Syntax
SENSe:SWEep:OFFSet:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR