Agilent Technologies N3280A TV Cables User Manual


 
6 – Language Dictionary
68
[SOURce:]VOLTage:MODE
This command determines what happens to the output voltage during a triggered event.
FIXed
The output voltage is unaffected when a trigger occurs.
STEP
The output voltage is programmed to the value set by VOLT:TRIG when a trigger occurs.
Command Syntax
[SOURce:]VOLTage:MODE <mode>, (@<channel list>)
Parameters
FIXed | STEP
*RST Value
FIXed
Examples
VOLT:MODE FIX,(@1)
Query Syntax
[SOURce:]VOLTage:MODE? (@<channel list>)
Returned Parameters
<CRD>
Related Commands
VOLT:TRIG
[SOURce:]VOLTage:PROTection:STATe
This command enables or disables the overvoltage protection (OVP) function. The command only
applies in voltage priority mode. When enabled, the output of the unit will shut down and the output
relays will open when the output voltage exceeds +11.5V (±0.3V), or 11.5V (±0.3V).
CAUTION: Disabling the overvoltage protection function may cause excessive output voltages, such
as can occur if remote sense leads are shorted, to damage the equipment under test.
Command Syntax
[SOURce:]VOLTage:PROTection:STATe <bool>, (@<channel list>)
Parameters
0 | OFF | 1 | ON
*RST Value
ON
Examples
VOLT:PROT:STAT 0,(@1)
Query Syntax
[SOURce:]VOLTage:PROTection:STATe? (@<channel list>)
Returned Parameters
<NR1>0 or 1