Agilent Technologies N3280A TV Cables User Manual


 
B - Performance and Calibration Procedures
98
Performance Test Record Form
Model Agilent N3280A - Output 1 Report No ______________ Date __________________
Test Description Minimum
Specification
Results Maximum
Specification
VOLTAGE PRIORITY TESTS
Programming Accuracy (DMM readings)
Voltage ( 0V)
Voltage (+10V)
Voltage (-10V)
+ 1mA Current limit
+ 0.5A Current limit
1mA Current limit
0.5A Current limit
2mV
9.988 V
9.988 V
0.949mA
0.49945 A
0.949mA
0.49945 A
_________
_________
_________
_________
_________
_________
_________
+ 2mV
10.012 V
10.012 V
1.051mA
0.50055 A
1.051mA
0.50055 A
Readback Accuracy (MEAS? readings)
Voltage ( 0V)
Voltage (+10V)
Voltage (-10V)
+ 0.5A range current
0.5A range current
+ 15mA range current
15mA range current
+ 0.5mA range current
0.5mA range current
2mV
Vout
12mV
Vout
12mV
Iout
0.7mA
Iout
0.7mA
Iout
15µA
Iout
15µA
Iout
0.7µA
Iout
0.7µA
_________
_________
_________
_________
_________
_________
_________
_________
_________
+ 2mV
Vout + 12mV
Vout + 12mV
Iout + 0.7mA
Iout + 0.7mA
Iout + 15
µA
Iout + 15
µA
Iout + 0.7
µA
Iout + 0.7
µA
Load Effect
Voltage
+ Current limit
Current limit
400µV
30µA
30µA
_________
_________
_________
+ 400
µV
+ 30
µA
+ 30µA
Source Effect
Voltage
+ Current limit
Current limit
200µV
10µA
10µA
_________
_________
_________
200µV
10µA
10µA
PARD (Ripple and Noise)
Voltage (rms)
Voltage (peak-to-peak)
±Current limit (rms)
_________
_________
_________
380
µV
4mV
40
µA
Transient Response Time Low( 10kHz):
Med ( 20kHz):
High ( 30kHz):
_________
_________
_________
60
µs
45µs
35µs
CURRENT PRIORITY TESTS
Programming Accuracy (DMM readings)
Current ( 0A )
Current ( 0.5mA)
Current (
0.5mA)
1µA
0.0004985 A
0.0004985 A
_________
_________
_________
+ 1µA
0.0005015 A
0.0005015 A
Load Effect
Current
25nA
_________ + 25nA
Source Effect
Current
10nA
_________ + 10nA
PARD (Ripple and Noise)
Current (rms)
_________
1.5
µA
Transient Response Time
_________
90µs