Agilent Technologies N3280A TV Cables User Manual


 
Language Dictionary - 6
67
[SOURce:]FUNCtion:MODE
This comand configures the output operating mode. Note that if the output is on, changing the output
mode will cause the output to cycle OFF, then ON.
VOLTage
Configures the output for voltage priority operation
CURRent
Configures the output for current priority operation
Command Syntax
[SOURce:]FUNCtion:MODE <mode>,(@<channel list>)
Parameters
VOLTage | CURRent
*RST Value
VOLT
Examples
FUNC:MODE VOLT,(@1)
Query Syntax
[SOURce:]FUNC:MODE? (@<channel list>)
Returned Parameters
<CRD>
[SOURce:]VOLTage:ALC:BWIDth
This command configures the output compensation band for the voltage mode. If capacitive loads cause
the output to oscillate, use this command to select a lower compenstion band. Note that if the output is
on, changing the compensation will cause the output to cycle OFF, then ON. The following
compensation bandwidths may be programmed: 30 kHz, 20 kHz, or 10 kHz.
Command Syntax
[SOURce:]VOLTage:ALC:BWIDth <Nrf>, (@<channel list>)
Parameters
10000 | 20000 | 30000
*RST Value
30000
Examples
VOLT:ALC:BWID 10000,(@1)
Query Syntax
[SOURce:]VOLTage:ALC:BWIDth? (@<channel list>)
Returned Parameters
<NR3>
[SOURce:]VOLTage[:IMMediate]
[SOURce:]VOLTage:TRIGgered
These commands set the immediate and the pending triggered voltage level of the dc source. The
immediate level is the voltage programmed for the output terminals. The pending triggered level is a
stored value that is transferred to the output terminals when a trigger occurs. To respond to a trigger, the
[SOUR:]VOLT:MODE must be set to STEP, and the trigger system must be initiated.
Command Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]<NRf+>, (@<channel list>)
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <Nrf+>, (@<channel list>)
Parameters
see table 6-3
Default Suffix
V (volts)
*RST Value
0
Examples
VOLT 2.5,(@1) VOLT:TRIG 20,(@1)
Query Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? (@<channel list>)
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? (@<channel list>)
Returned Parameters
<NR3>
Related Commands
INIT VOLT:MODE