Agilent Technologies N3280A TV Cables User Manual


 
Language Dictionary - 6
63
SENSe:SWEep:POINts
This command defines the number of points in a measurement.
Command Syntax
SENSe:SWEep:POINts <NRf+>
Parameters
1 through 4096
*RST Value
5
Examples
SENS:SWE:POIN 1024
Query Syntax
SENSe:SWEep:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR
SENSe:SWEep:TINTerval
This command defines the time period between samples. The value that you enter for the time interval
will be rounded to the nearest 30.4 microsecond increment.
Command Syntax
SENSe:SWEep:TINTerval <NRf+>
Parameters
30.4 microseconds through 60800 seconds
*RST Value
30.4 microseconds
Examples
SENS:SWE:TINT 60.8E-6
Query Syntax
SENSe:SWEep:TINTerval?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR
SENSe:WINDow
This command sets the window function that is used in dc measurement calculations. The following
functions can be selected:
HANNing
A signal conditioning window that reduces errors in dc measurement calculations in
the presence of periodic signals such ac line ripple. The Hanning window multiplies
each point in the measurement sample by the function cosine
4
.
RECTangular
A window that returns measurement calculations without any signal conditioning.
NOTE: Neither window function alters the instantaneous voltage or current data returned in the
measurement array.
Command Syntax
SENSe:WINDow[:TYPE] <type>
Parameters
HANNing | RECTangular
*RST Value
RECTangular
Examples
SENS:WIND RECT
Query Syntax
SENSe:WINDow[:TYPE]?
Returned Parameters
<CRD>