Remote Operation
Common Commands
*OPC
Sets bit 0 in the Standard Event Status register when all pending operations
have finished.
*OPC?
Returns a 1 when all operations have finished.
*OPT?
Returns a comma-separated list of strings that identifies the optical spectrum
analyzer's option configuration. A 0 indicates no options are present.
*RCL <numeric_value>|<filename>[,INTernal|FLOPpy]
Recalls previously saved instrument settings from the requested register or
file.
*RST
Executes a device reset and returns the instrument to a known state. This
command is not the same as the SYSTem:PRESet command. See page 3-5 for a
list of the instrument preset conditions. The *RST settings are the same as for
SYSTem:PRESet, except that Repeat Sweep is turned
OFF
.
*SAV <numeric_value>|<filename>[,INTernal|FLOPpy]
Saves instrument settings to the designated register or file.
*SRE <numeric_value>
Sets the bits in the Service Request Enable register. The parameter is rounded
to an integer value and interpreted as a binary number, representing the bit
values of the register. The Service Request Enable register serves as a mask
for the Status Byte. When a bit in the Status Byte goes to 1, if the correspond-
ing bit in the Service Request Enable register is a 1, the instrument asserts the
Service Request line on the HP-IB.
*SRE?
Returns the value of the Service Request Enable register.