HP (Hewlett-Packard) 86140A TV Cables User Manual


 
6-14
Reference
Cleaning Connections for Accurate Measurements
Repeated connections made without removing loose particles, or using
improper cleaning tools can lead to physical damage of the glass fiber endface,
as shown in Figure 6-4. When severe, the damage on one connector end can
be transferred to another good connector that comes in contact with it.
The cure for these problems is disciplined connector care. as described in the
following list and in Cleaning Connections for Accurate Measurements on
page 6-10.
Figure 6-4. A dirty endface from poor cleaning
Figure 6-5. A damaged fiber end from using an improper cleaning procedure
book.book Page 14 Monday, January 31, 2000 10:34 AM