VT1422A Command Reference 263Chapter 7
CALibration
Figure 7-3. Levels of Working Calibration
Subsystem Syntax CALibration
:CONFigure
:RESistance
:VOLTage <range>, ZERO | FS
:REMote?
:DATA <cal_data_block>
:DATA?
:STORe
:SETup
:SETup?
:STORe ADC | TARE
:TARE (@<ch_list>)
:RESet
:TARE?
:VALue
:RESistance <ref_ohms>
:VOLTage <ref_volts>
:ZERO?
CALibration:CONFigure:RESistance
CALibration:CONFigure:RESistance connects the on-board reference resistor to
the Calibration Bus. A four-wire measurement of the resistor can be made with an
external multimeter connected to the HCAL, LCAL, HOHM and LOHM terminals
on the Terminal Module or the VH, VL, Ω H, and Ω L terminals on the Cal Bus
connector when not using a terminal module.
SCP
SCP
SCP
SCP
SCP
8
8
8
8
8
8
8
8
SCP
SCP
SCP
Multiplexer
CAL:ZERO?
CAL:TARE
*CAL? or
CAL:SETup
User’s System Wiring