VXI VT1422A Universal Remote User Manual


 
VT1422A Command Reference 339Chapter 7
[SENSe]
[SENSe:]FREQuency:APERture?
[SENSe:]FREQuency:APERture? <channel> returns the frequency counting
gate time.
Parameters
Comments If the channel specified is on an SCP that doesn't support this function, an error
will be generated. See the SCP's User's Manual for its capabilities.
Related Commands: SENSe:FREQuency:APERture
Returned Value: Returns numeric gate time in seconds. The type is float32.
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
[SENSe:]FUNCtion:CONDition
[SENSe:]FUNCtion:CONDition <ch_list> sets the SENSe function to input the
digital state for channels in <ch_list>. Also configures digital SCP channels as inputs
(this is the *RST condition for all digital I/O channels).
Parameters
Comments The VT1533A SCP senses 8 digital bits on each channel specified by this
command. The VT1534A SCP senses 1 digital bit on each channel specified by
this command.
If the channels specified are not on a digital SCP, an error will be generated.
Use the INPut:POLarity command to set input logical sense.
Related Commands: INPut:POLarity
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital
SCP channels.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage To set second 8-bits of VT1533A at SCP position 4 and upper 4-bits of VT1534A at
SCP position 5 to digital inputs send:
SENS:FUNC:COND (@133,144:147)
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel string 100 - 163 none
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
ch_list string 100 - 163 none