344 VT1422A Command Reference Chapter 7
[SENSe]
Usage program must put table constants into array table_block
DIAG:CUST:PIEC 1,table_block,(@100:107)
send characterized thermocouple table
for use by channels 0-7
SENS:FUNC:CUST:TC N,.25,(@100:107)
link custom thermocouple EU with chs
0-7, use reference temperature
compensation for N type wire.
SENSE:REF RTD,92,(@120)
designate a channel to measure the
reference junction temperature
include these channels in a scan list (REF channel first)
INITiate then TRIGger module
[SENSe:]FUNCtion:FREQuency
[SENSe:]FUNCtion:FREQuency (@<ch_list>) sets the SENSe function to
frequency for channels in <ch_list>. Also configures the channels specified as digital
inputs.
Parameters
Comments • If the channels specified are on an SCP that doesn't support this function,
an error will be generated. See the SCP's User's Manual for its capabilities.
• Use the SENSe:FREQuency:APERture command to set the gate time for the
frequency measurement.
• Related commands: SENS:FREQ:APER
• *RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital
SCP channels
• Send with VXIplug&play Function: hpe1422_cmd(...)
Usage SENS:FUNC:FREQ (@144)
set channel 44's to frequency
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
ch_list string 100 - 163 none