Teledyne API Model 100AH SO
2
Analyzer Instruction Manual, 02417, Rev. D
2 GETTING STARTED
2.1 Installation
CAUTION
To avoid personal injury, always use two persons to
lift and carry the Model 100AH.
1. Verify that there is no apparent shipping damage. If damage has occurred please advise
shipper first, then Teledyne API.
2. Before operation it is necessary to remove the shipping hold-down screws. Remove the
instrument cover, then refer to Figure 2-1 for screw location.
NOTE
Save these shipping screws and re-install then whenever the
unit is shipped to another location.
3. While the instrument cover is removed, please check the voltage and frequency label on the
cover of the power supply module and compare that to your local power before plugging in
the M100AH.
4. Check for internal shipping damage, and generally inspect the interior of the instrument to
make sure all circuit boards and other components are in good shape.
5. Replace the instrument cover.
6. When installing the M100AH, allow at least 4” (100 mm) clearance at the back and at least
1” (25 mm) clearance at each side for proper venting.
2-1