Teledyne API Model 100AH SO
2
Analyzer Instruction Manual, 02417, Rev. D
Table 2-1: Final Test and Calibration Values (Continued)
Parameter
Observed
Value
Units Nominal Range Reference Section
Span and Cal Values
SO
2
Span Conc PPM 100 - 4500 Table 8-3
SO
2
Slope
1.0
±
0.3
8.7
SO
2
Offset mV <100 8.7
Noise at Zero
(rms)
PPM 0.05 - 0.2 Table 10-1
Noise at Span
(rms)
PPM 0.5% of reading
(above 50PPM)
Table 10-1
Measured Flows
Sample Flow cc/min
650
±
60
10.2.2, 10.3.5,
Figure 10-5
Factory Installed Options Option Installed
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Zero/Span Valves Manifold
Multi-drop
Kicker
37 mm Filter
4-20 mA Isolated Current Loop Output
(non-isolated standard)
ISOLATED NON-ISOLATED
SO
2
(RANGE 1) ____________ ____________
SO
2
(RANGE 2) ____________ ____________
SPARE ____________ ____________
TEST OUTPUT ____________ ____________
PROM # Serial #
Date Technician
2-13