Analyzer Instruction Manual, 02417, Rev. D
Example 3: To perform zero-span calibration check once per day at 10:30 PM and zero calibration
adjust once per week starting at 11:30 PM, 5/20/97.
1. Select any one of SEQx and setup as example 1 above.
2. Select any other SEQx and program as follows. Always avoid setting two or more sequences
at the same time of the day.
1) MODE: ZERO
2) TIMER ENABLE: ON
3) STARTING DATE: 5/20/97
4) STARTING TIME: 23:30
5) DELTA DAYS: 7
6) DELTA TIME: 00:00
7) DURATION: 15.0 MINUTES
8) CALIBRATE: ON