Teledyne API Model 100AH SO
2
Analyzer Instruction Manual, 02417, Rev. D
The data in the default Data Channels can be viewed through the SETUP-DAS-VIEW menu.
Use the PREV and NEXT buttons to scroll through the Data Channels and press VIEW to view
the data. The last record in the Data Channel is shown. Pressing PREV and NEXT will scroll
through the records one at a time. Pressing NX10 and PV10 will move forward or backward 10
records. For Data Channels that log more than one parameter, such as PNUMTC, buttons labeled
<PRM and PRM> will appear. These buttons are used to scroll through the parameters located
in each record.
The function of each of the default Data Channels is described below:
CONC:
Samples SO
2
concentration at one minute intervals and stores an average every
hour with a time and date stamp. Readings during calibration and calibration
hold off are not included in the data. The last 800 hourly averages are stored.
PNUMTC:
Collects sample flow and sample pressure data at five minute intervals and
stores an average once a day with a time and date stamp. This data is useful for
monitoring the condition of the pump and critical flow orifice (sample flow)
and the sample filter (clogging indicated by a drop in sample pressure) over
time to predict when maintenance will be required. The last 360 daily averages
(about 1 year) are stored.
CALDAT:
Logs new slope and offset every time a zero or span calibration is performed.
This Data Channel also records the instrument reading just prior to performing
a calibration.
NOTE: This Data Channel collects data based on an event (a calibration) rather
than a timer. This Data Channel will store data from the last 200 calibrations.
This does not represent any specific length of time since it is dependent on how
often calibrations are performed. As with all Data Channels, a time and date
stamp is recorded for every data point logged.
5-13