5.3.3.2 RS-232 Reporting
Automatic RS-232 reporting can be independently enabled and disabled for each Data Channel.
For all default data channels, RS-232 reporting is initially set to “OFF.” If this property is turned
on, the Data Channel will issue a report with a time and date stamp to the RS-232 port every
time a data point is logged. The report format is shown below:
D 31:10:06 0412 CONC : AVG CONC1=6.8 PPM
The report consists of the letter “D” follow by a time/date stamp (“31:10:06”) followed by the
instrument ID number (“0412”). Next is the Data Channel name (“CONC”) and the sampling
mode (“AVG” indicates that the data point is an average of more than one sample as opposed to
a instantaneous reading, “INST”). Finally, the name of the parameter and its value
(“CONC1=6.8 PPM”) are printed. For Data Channels that sample more than one parameter, such
as PNUMTC and CALDAT, each parameter is printed on a separate line.
To enable RS-232 reporting for a specific Data Channel:
Step Action Comment
1.
Press SETUP-DAS-
EDIT-ENTR
Enter DAS menu to edit Data Channels
2.
Press PREV/NEXT
Select Data Channel to edit
3.
Press EDIT
Edit selected Data Channel
4.
Press SET> (5 times)
Scroll through setup properties until RS-232 REPORT:
OFF is displayed
5.
Press EDIT
Edit selected setup property
6.
Toggle OFF to ON
Change RS-232 REPORT property
7.
Press ENTR
Accepts change
8.
Press EXIT (4 times)
Exits back to sample menu
See Section 7.7 for more information on DAS reporting through the RS-232 interface.
5-14