Analyzer Instruction Manual, 02417, Rev. D
3 SPECIFICATIONS, AGENCY APPROVALS,
WARRANTY
3.1 Specifications
Ranges In 1 PPM increments from 10 PPM to 5000 PPM, dual ranges or autoranging
Noise at Zero 0.05 PPM RMS
Noise at Span <0.5% of reading (above 50 PPM)
Lower Detectable Limit
1
0.1 PPM RMS
Zero Drift
2
< 1 PPM/24 hours
Zero Drift
2
<2 PPM/7 days
Span Drift
2
<0.5% FS/7 days
Lag Time 5 sec
Rise Time 95% in < 30 sec
Fall Time 95% in < 30 sec
Sample Flow Rate
650cc/min.
±
10%
Linearity 1% of full scale
Precision 0.5% of reading
Temperature Range 5 - 40
o
C
Temp Coefficient < 0.1% per
o
C
Humidity 0 - 95% RH, non-condensing
Voltage Coefficient < 0.05% per V
Dimensions HxWxD 7"x17"x23.6" (18 cm x 43 cm x 61 cm)
Weight, Analyzer 43 lbs (19.5 kg)
Weight, Pump Pack 16 lbs (7 kg)
Power, Analyzer
110 v
∼
60 Hz, 220 v
∼
50 Hz, 240 v
∼
50 Hz, 250 watts
Power, Analyzer
3
230 v
∼
50 Hz, 2.5A
Power, Ext. Pump
110 v
∼
60 Hz, 220 v
∼
50 Hz, 240 v
∼
50 Hz, 295 watts
Power, Ext. Pump
3
230 v
∼
50 Hz, 2.5 A
Environmental
Conditions
Installation Category (Overvoltage Category) II
Pollution Degree 2
Recorder Output
4
4 - 20 mA non-isolated standard, 0-100 mV, 0-1, 5, 10 v ; resolution of 1 part
in 1024 of selected voltage or current range. 4 - 20 mA isolated optional.
Status Option 12 Status Outputs from opto-isolator
Measurement Units PPM, mg/m
3
1. Defined as twice the zero noise level.
2. At constant temperature and voltage.
3. Electrical ratings for CE Mark compliance.
4. Bi-polar. (voltage or current selectable by the jumper on the motherboard)