Teledyne 100AH TV Converter Box User Manual


 
Teledyne API Model 100AH SO
2
Analyzer Instruction Manual, 02417, Rev. D
APPENDIX A MAINTENANCE SCHEDULE FOR M100AH
DATE INSTRUMENT RECEIVED _____________
JAN FEB MAR APR MAY JUNE JULY AUG SEPT OCT NOV DEC
PARTICULATE
FILTER ELEMENT
REPLACE WEEKLY AS NEEDED
PUMP CHARCOAL
SCRUBBER
REPLACE QUARTERLY
DFU
PARTICULATE
FILTER FOR
SCRUBBER
REPLACE ANNUALLY
PNEUMATIC
LINES
EXAMINE AND REPLACE AS
NECESSARY
LEAK CHECK
AND QUICK CAL
CHECK ANNUALLY. SEE
SECTION IN MANUAL. LEAK
CHECK AFTER ANY
MAINTENANCE THEN FOLLOW
QUICK CAL.
PUMP
DIAPHRAGMS
EVERY 12 MONTHS
SAMPLE ORIFICE REPLACE O-RINGS AND FILTER
ANNUALLY
SAMPLE FLOW CHECK FOR PROPER FLOW
650 ± 10% CC/MIN. ANNUALLY
A-1