Analyzer Instruction Manual, 02417, Rev. D
Table 7-5: RS-232 Command Summary
Commands Definition
? Print help screen
T SET ALL Enable display of all test variables during T LIST
T SET name Display only NAME during T LIST
T LIST [ALL] Print all test variables enabled with T SET or ALL
warnings
T name Print single test, "name" from Table 7-7
T CLEAR ALL Disable T LIST, use with T SET name
W SET ALL Enable display of all warnings during W LIST
W LIST [ALL] Print warnings enabled with W SET or ALL warnings
W name Print individual "name" warning from Table 7-8
W CLEAR ALL Disable W LIST, use with W SET
C command Execute calibration "command" from Table 7-10
D LIST Prints all I/O signal values
D name Prints single I/O signal value/state
D name=value Sets variable to new "value"
D LIST NAMES Lists diagnostic test names
D ENTER name Enters and starts 'name' diagnostic test
D EXIT Exits diagnostic mode
D RESET Resets analyzer (same as power-on)
D RESET RAM System reset, plus erases RAM. Initializes DAS, SO
2
concentration readings, calibration not affected.
D RESET EEPROM System reset, plus erases EEPROM (RAM_RESET
actions + setup variables, calibration to default values)
V LIST Print all easy variable names from Table 10-5
V name Print individual "name" variable
V name=value Sets variable to new "value"
V CONFIG Print analyzer configuration
7-6