Teledyne API Model 100AH SO
2
Analyzer Instruction Manual, 02417, Rev. D
Table 2-1: Final Test and Calibration Values
Test Values
Observed
Value
Units Nominal Range Reference Section
RANGE PPM 10 - 5000 5.3.4
STABIL PPM 0.01 - 0.05 5.2.1, 10.1.1,
10.2.5, Table 10-1
INLET
SAMPLE
PRESS
in-Hg-
Abs
25 - 35 10.1.1, 10.3.5,
Table 10-1
VACUUM
PRESS
in-Hg-
Abs
4 - 10 10.3.5, Table 10-1
SAMP FL CC / MIN
650
±
60
10.2.2, 10.3.5,
Table 10-1
PMT mV 0 - 5000 10.4.1
UV LAMP mV 2000 - 4000
typical
10.4.2
STR. LGT PPM <100.0 Table 10-1
DARK PMT mV <200 Table 10-1
DARK LAMP mV <200 Table 10-1
SLOPE
1.0
±
0.3
8.7
OFFSET mV <200 8.7
HVPS V 450 - 900 constant 10.3.9
DCPS mV
2500
±
200
10.3.4
RCELL TEMP
o
C
50
±
1
10.3.6
BOX TEMP
o
C 8 - 50 10.3.3.1
PMT TEMP
o
C
7
±
1
10.3.8
Electric Test & Optic Test
Electric Test
PMT Volts mV
2000
±
100
10.1.3.2
SO
2
Conc PPM
2000
±
100
10.1.3.2
Optic Test
PMT Volts mV
200
±
20
10.1.3.3
SO
2
Conc PPM
200
±
20
10.1.3.3
2-12