9 MAINTENANCE
NOTE
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
9.1 Maintenance Schedule
Table 9-1: Preventative Maintenance Schedule
Item Maintenance Interval Reference Section
TEST functions Check every month Table 10-1
Zero/Span Calibration Weekly or as needed Section 8
Zero/Span checks Daily or a needed Section 8, Table 6-2
Particulate Filter Weekly as needed Figure 9-1
Pump charcoal scrubber Refill every 3 months Section 9.3
Reaction cell cleaning Clean annually or as necessary Section 9.5, Figure 9-2
Sample Flow Check every 6 months Figure 9-3, Section 10.2.2
Pneumatic Lines Examine every 12 months,
clean if necessary
Figure 9-3
Factory Calibration Calibrate each year or after
repairs
Section 10.1.6
Leak Check Check every 6 months Section 9.7