Teledyne 102E TV Converter Box User Manual


 
GETTING STARTED M102E/M501 TRS
(Addendum to M101E Manual - P/N 04740 Rev A)
Source of
SAMPLE Gas
VENT if input is pressurized
MODEL 700
Gas Dilution Calibrator
(with O
3
generator option)
VENT
Calibrated
H
2
S gas
(At high
concentration)
MODEL
102E
From Converter
Sam
p
le
Zero Air
Exhaust
M501-TRS
To Converter
From Analyze
r
To Analyze
r
MODEL 701
Zero Air
Generator
External Zero
Air Scrubber
Filter
Figure 3-10: Pneumatic Connections for Formal Calibration of Units with an IZS Valve
Option
Source of
SAMPLE Gas
VENT if input is pressurized
Ambient
Air
MODEL
102E
From Converter
Sam
p
le
Zero Air
(
Scrubber
)
Exhaust
S
p
an
M501-TRS
To Converter
From Analyze
r
To Analyzer
Figure 3-11: Pneumatic Connections for Informal Calibration Checks of Units with IZS
Valve Option
NOTE
Gas flow must be maintained at all times for units with IZS Options installed. The IZS
option requires a permeation tube ( customer supplied ) which emits H
2
S. Insufficient
gas flow can build up H
2
S to levels that will damage the instrument.
Remove the permeation device when taking the analyzer out of operation.
16 05514 Rev A1