Teledyne 102E TV Converter Box User Manual


 
M102E/M501 TRS SPECIFICATIONS, APPROVALS AND WARRANTY
(Addendum to M101E Manual - P/N 04740 Rev A)
2. SPECIFICATIONS, APPROVALS AND
WARRANTY
2.1. Specifications
There are no significant differences between the performance specifications for the M102E and the
M102E as listed in Section 2.1 of the M101E Manual - P/N 04740 Rev A.
2.1.1. M501-TRS Specifications
Table 2-1: Model 102E Basic Unit Specifications
Minimum Converter Efficiency
H
2
S >95%
COS >90%
CS
2
>90%
Maximum TS Concentration for
specified conversion efficiency
20 ppmv
Sample Flow Rate 650cc/min. ±10% - driven by M102E pneumatic system
Optimum Converter
Temperature
1000°C (factory setup)
Maximum Converter
Temperature
1100°C
Dimensions H x W x D 7" x 17" x 22" (178 mm x 432 mm x 559 mm)
Weight
16 lbs (8 kg)
26 lbs (12 kg) CE version
AC Power Rating
115 V, 50/60 Hz - 400 Watts;
230 V, 50/60 Hz - 575 Watts; CE Version
Internal Alarms
High Alarm Point: 1050°C
Low Alarm Point: 950°C
Alarm Output Relay
SPST - 1 point: Alarm output is energized should either the temperature
controller’s high or low internal alarm set points be activated.
Alarm Output Rating
220V AC/30V DC, 1A (resistive load)
Environmental Installation category (over-voltage category) II; Pollution degree 2
Certifications
IEC 1010-1 / 61010-1:93 (includes A1) + A2:95,
For indoor use at altitudes 2000m only
2.2. EPA Equivalency Designation
No EPA equivalency standards exist for TRS measurement, however, the M102E analyzer qualifies
for EPA equivalency designation as Reference Method Number EQSA-0495-100 per 40 CFR Part 53
when operated under the following conditions:
Measurement Mode: SO2 single gas mode.
Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
Ambient temperature range of 5
o
C to 40
o
C.
Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
05514 Rev A1 3