Teledyne 102E TV Converter Box User Manual


 
APPENDIX A-2: Setup Variables For Serial I/O, Revision A.2 M102E/M501 TRS
(Addendum to M101E Manual - P/N 04740 Rev A)
APPENDIX A-2: Setup Variables For Serial I/O, Revision A.2
NOTE
Setup Variables for the M102E are the same as those for the M102E (see Appendix A-3
of the M101E Manual - P/N 04740110 Rev A) with the following exceptions:
Table A-1: Changed or added Setup Variables for M102E Software Revision A.2
SETUP VARIABLE NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
MEASURE_MODE SO2-TRS SO2, SO2-TRS,
TRS
Gas measurement mode. Enclose
value in double quotes (") when
setting from the RS-232 interface.
TRS_SPAN1 Conc 400 0.1–50000 Target TRS concentration during
span calibration of range 1.
TRS_SPAN2 Conc 400 0.1–50000 Target TRS concentration during
span calibration of range 2.
TRS_SLOPE1 PPB/mV 1 0.25–4 TRS slope for range 1.
TRS_SLOPE2 PPB/mV 1 0.25–4 TRS slope for range 2.
TRS_OFFSET1 mV 0 -1500–1500 TRS offset for range 1.
TRS_OFFSET2 mV 0 -1500–1500 TRS offset for range 2.
CE_FACTOR1 1 0.8–1.2 Converter efficiency factor for TRS
for range 1.
CE_FACTOR2 1 0.8–1.2 Converter efficiency factor for TRS
for range 2.
REMOTE_CAL_MODE — SO2-
LOW
SO2-LOW,
SO2-HIGH,
TRS-LOW,
TRS-HIGH
Gas and range to calibrate during
contact-closure and Hessen
calibration. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
STABIL_GAS — SO2 SO2,
TRS
Gas to use to measure concentration
stability. Enclose value in double
quotes (") when setting from the RS-
232 interface.
Table A-2: Deleted Setup Variables for M102E Software Revision A.2
SETUP VARIABLE NUMERIC
UNITS
DEFAULT VALUE VALUE
RANGE
DESCRIPTION
CONV_TYPE
2
— MOLY NONE,
MOLY
Converter type.
CONV_SET
2
ºC 315
Warnings:310–320
0–350 Converter temperature set point and
warning limits.
TEST_CHAN_ID — NONE CONV
TEMP
,
All other Test Channel settings are
the same as those listed in appendix
A-1 of the M101E Manual - P/N
04740110 Rev A
A-6 05515 Rev A.2