Teledyne 102E TV Converter Box User Manual


 
M102E/M501 TRS APPENDIX A-3: Warnings and Test Functions, Revision A.2
(Addendum to M101E Manual - P/N 04740 Rev A)
APPENDIX A-3: Warnings and Test Functions, Revision A.2
NOTE
Warning messages and test functions for the M102E are the same as those for the
M102E (see Appendix A-3 of the M101E Manual - P/N 04740110 Rev A) with the
following exceptions:
Table A-4: Warning Messages deleted from M102 Software Revision A.2
NAME MESSAGE TEXT DESCRIPTION
WCONVTEMP CONV TEMP WARNING Converter temperature outside of warning limits specified
by CONV_SET variable.
Table A-5: Test Functions Changed and Added to M102 Software Revision A.2
TEST Function Message Text DESCRIPTION
STABILITY TRS STB
1
=0.0 PPB
1
Concentration stability (standard deviation based on
setting of STABIL_FREQ and STABIL_SAMPLES).
SO
2
SLOPE SO2 SLOPE
2
=1.061 Slope for current SO
2
measurement range, computed
during zero/span calibration.
SO
2
OFFSET SO2 OFFS
2
=250.0 MV Offset for current TRS measurement range, computed
during zero/span calibration.
TRS SLOPE TRS SLOPE
2
=1.061 Slope for current SO
2
measurement range, computed
during zero/span calibration.
TRS OFFSET TRS OFFS
2
=250.0 MV Offset for current TRS measurement range, computed
during zero/span calibration.
1
Shown as it appear when analyzer is in TRS mode. In SO
2
mode appear as SO2 STB. In multigas
mode, both versions appear.
2
Only Appears in SO
2
and multigas measurement modes.
2
Only Appears in TRS and multigas measurement mode.
Table A-6: Test Functions Deleted from M102 Software Revision A.2
TEST Function Message Text DESCRIPTION
RESPONSE
2
RSP=1.11(0.00) SEC Instrument response. Length of each signal processing
loop. Time in parenthesis is standard deviation.
VACUUM
5
VAC=9.1 IN-HG-A Vacuum pressure.
RCELLDUTY RCELL ON=0.00 SEC Sample chamber temperature control duty cycle.
IZSDUTY IZS ON=0.00 SEC IZS temperature control duty cycle.
CONVTEMP CONV TEMP=315.0 C TRS Æ SO
2
Converter temperature.
05515 Rev A.2 A-7