3.8. Initial Calibration
Initial calibration of the M102E should be performed with:
• Zero air supplied by a zero air generator such as the Teledyne Instruments’ M701;
• Calibrated H
2
S span gas of the appropriate concentration:
• With external pneumatic connections as described in Figures 3-7 through 3-11 of this
addendum, and;
• Using the information and procedure included in Section 3.3 of the M101E Manual - P/N
04740 Rev A.
No initial calibration of the M501-TRS temperature controller is required.
NOTE
Once you have completed the above set-up procedures, please fill out the quality
questionnaire that was shipped with your unit and return it to Teledyne Instruments.
This information is vital to our efforts in continuously improving our service and our
products. Thank you.