Agilent Technologies 1660AS TV Converter Box User Manual


 
Query :RTC?
The RTC query returns the real-time clock setting.
Returned Format
[:RTC] <day>,<month>,<year>,<hour>,<minute>,<second>
Example OUTPUT XXX;":RTC?"
SELect
Command :SELect <module>
The SELect command selects which module (or system) will have parser
control. SELect defaults to System (0) at power up. The appropriate module
(or system) must be selected before any module (or system) specific
commands can be sent. SELECT 0 selects the System, SELECT 1 selects the
logic analyzer (state and timing), and SELECT 2 selects the oscilloscope
module. Select –2, –1 and, 3 through 10 are accepted but no action will be
taken. When a module is selected, the parser recognizes the module’s
commands and the System/Intermodule commands. When SELECT 0 is
used, only the System/Intermodule commands are recognized by the parser.
Figure 9-2 shows the command tree for the SELect command.
The command parser in the 1660-series logic analyzers is designed to accept
programs written for the 16500A logic analysis system with a 16550A logic
analyzer module; however, if the parameters 3 through 10 are sent, the
1660-series logic analyzer will take no action.
<module> An integer 0 through 2 (–2, –1, and 3 through 10 unused).
Example OUTPUT XXX;":SELECT 0"
Mainframe Commands
SELect
9–20