Agilent Technologies 1660AS TV Converter Box User Manual


 
SEQuence
Command :MACHine{1|2}:STRigger:SEQuence
<number_of_levels>,
<level_of_trigger>
The SEQuence command redefines the state analyzer trace sequence. First,
it deletes the current trace sequence. Then it inserts the number of levels
specified, with default settings, and assigns the trigger to be at a specified
sequence level. The number of levels can be between 2 and 12 when the
analyzer is armed by the RUN key.
<number_of_
levels>
An integer from 2 to 12
<level_of_
trigger>
An integer from 1 to (number of existing sequence levels 1)
Example OUTPUT XXX;":MACHINE1:STRIGGER:SEQUENCE 4,3"
Query :MACHine{1|2}:STRigger:SEQuence?
The SEQuence query returns the current sequence specification.
Returned Format
[:MACHine{1|2}:STRigger:SEQuence] <number_of_levels>,
<level_of_trigger><NL>
<number_of_
levels>
An integer from 2 to 12
<level_of_
trigger>
An integer from 1 to (number of existing sequence levels 1)
Example OUTPUT XXX;":MACHINE1:STRIGGER:SEQUENCE?"
STRigger (STRace) Subsystem
SEQuence
16–16